Currency RUB
Analyzer of a range Agilent Technologies E4407B buy in Moscow
Buy Analyzer of a range Agilent Technologies E4407B
Analyzer of a range Agilent Technologies E4407B

Analyzer of a range Agilent Technologies E4407B

In stock
Reconfirm the price with seller
Shipping:
Russia, Moscow
(View map)
Technical characteristics
  • BrandAgilent Technologies
  • Country of manufactureUnited States
Description

Analyzer of a range Agilent Technologies E4407B

E4407B Agilent Technologies range analyzer Range of frequencies is from 9 kHz to 26,5ggts Permission strip from 100 Hz to 5 MHz The displayed average level of noise: - 150 dBm Dynamic range for harmonicas of the 3rd order: 108 dB (+15dbm TOI) General error of the level of 0,4 dB Attenuator with a step 5 dB The number of measurements with display on the display> 40izm/s Narrow-band digital filters from 1 Hz to 300 Hz (option) The watching generator (option) AM demodulator Procedures for measurement of ACPR, TOI, OBW, ACP, CCDF, distortion of harmonicas, SEM, etc. Options for measurement of a form of noise, the analysis of the modulated signals and phase noise Options for measurement of cdmaOne, W-CDMA, GSM/EDGE/GPRS, cdma2000, Cable TV, Bluetooth Built-in floppy-disk store Measurement of distance before damage to the VCh-cable (DTF) (option) LAN, RS-232 and GPIB interfaces Software of "IntuiLink"

Technical characteristics of the analyzer of a range of Agilent E4407B:

Parameters Values
Frequency measurements


Frequency range
Internal mixer 9 kHz to 26,5ggts
DC (Opt-UKB) 30 Of Hz to 26,5ggts
AC (Opt-UKB) 10 MHz to 26,5ggts
Strip N 0 1- 9 kHz to 3,0ggts
0 (Opt-UKB) 30gts6 to 3,0ggts
1 1- 2,85ggts to 6,7ggts
2 2- 6,2ggts to 13,2ggts
3 4- 12,8ggts to 19,2ggts
4 4- 18,7ggts to 26,5ggts
Source of basic frequency
Standard Optional (Opt-1D5)
Stability ±2 x 10-6/year ±1 x 10-7/year
Installation accuracy ±5 x 10-7 ±1 x 10-8
Temperature stability ±5 x 10-6 ±1 x 10-8 (at 20 … 30 °C)

Accuracy of reading of frequency

(Start, Stop, Center, Marker): ± (accuracy of reading x error of basic level chastoty1 + range accuracy +
+ 15% of a strip of permission + 10 Hz + 1 Hz x N)
Marker frequency counter Permission Chosen by the user: from 1 Hz to 100 kHz
Accuracy ± (frequency of a marker x error of basic level chastoty1 + resolution of the counter)
Review strip Range 0 Hz (a zero strip), from 100 Hz … Maximum frequency of the analyzer
Resolution 2 Of Hz x N
Accuracy (> 2000 points of development) ±0,5% of a strip of the review
Development time Strip width
> 0 Hz From 1 ms to 4000 with
= 0 Hz From 10 microsec to 4000 with
Opt - AYX From 50 nanoseconds to 4000 with
Opt - B7D From 25 nanoseconds to 4000 with
Accuracy ±1%
Development operation Free, single, linear, video detained, the external, displaced, temporary window (Opt-1D6), and on TV to a signal (Opt-B7B)
Operation delay range From 1 microsec to 400 with
Range of development From 101 to 8192 points
At a strip of the review of =0 Hz From 2 to 8192 points
Bandwidth
Range
From 1 kHz to 5 MHz (at-3 dB) at sequence 1-3-10
from 9 kHz to 120 kHz (at-6 dB) for EMI
Option - 1DR Adds Hz filters 10, 30, 100 and 300 (at-3 dB) and 200 Hz (at-6 dB) for EMI (for review strips 5mgts)

Accuracy

From 1 kHz to 3 MHz ±15%
5 MHz ±30%
From 1 Hz to 300 Hz (Opt-1DR)

±30%

Selectivity (characteristic)
- 60db/-3db

From 10 Hz to 300 Hz <5:16 digital, according to Gauss
From 1 kHz to 5 MHz <15:16 4 fields, according to Gauss are synchronously adjusted


Video a strip (-3 dB)
Width From 30 Hz to 3mgts6 at sequence 1-3-10
Option - 1DR Adds Hz filters 1, 3 and 10 (at bandwidth 1kgts)
Stability
Width of a radio strip (RBW) 1 of kHz, width of a video strip (VBW) of 30 Hz and detector of selections
Shift from CW on: Noise of a bokovoypolosa Typical value
≥1 kHz N of. . ≤ - 78dbts/Hz (-1D5)
≥10 kHz ≤ - 90dbts/Hz ≤ - 94dbts/Hz
≥20 kHz ≤ - 100dbts/Hz ≤ - 105dbts/Hz
≥30 kHz: ≤ - 106dbts/Hz ≤ - 112dbts/Hz
≥100 kHz: ≤ - 119dbts/Hz ≤ - 122dbts/Hz
≥1 MHz: ≤ - 125dbts/Hz ≤ - 127dbts/Hz
≥5 MHz: ≤ - 127dbts/Hz ≤ - 129dbts/Hz
≥10 MHz: ≤ - 131dbts/Hz ≤ - 136dbts/Hz
Option-120:
≥1 MHz: ≤ - 125dbts/Hz ≤ - 127dbts/Hz
≥5 MHz: ≤ - 127dbts/Hz ≤ - 129dbts/Hz
≥10 MHz: ≤ - 131dbts/Hz ≤ - 136dbts/Hz
Residual WC With a width of radio and video of strips of 1 kHz: scope (peak peak) of ≤150 Hz • N in 100 ms.
Option - 1D5 scope (peak peak) of ≤100 Hz • N4 in 100 ms.
Option - 1DR scope (peak peak) of ≤10 Gts6 • N4 in 20 ms.
Option - 1DR & - 1D5 scope (peak peak) of ≤2 Hz • N4 in 20 ms.
Lateral system strips Shift concerning CW signal 30kgts ≤ - 65dbts + 20 Log N
Amplitude measurements
Amplitude range Diapzon of measurements Displays the average level of noise to the maximum bezopasnomvkhodny level
Entrance range of an attenuator

From 0 to 65 dB at a step 5 dB

Detectors on routes Peak, on a negative pi-
ku, mean square on selections, averaging

The safest entrance level
Average constant power: (control of an entrance attenuator ≥15 dB) +30 dBm (1 W)
Peak pulse power: (control of an entrance attenuator ≥30 dB) +50 dBm (100 W)
DC 0Vdc.
Nonlinearity of the amplitude characteristic
1 dB
(the general power on the entrance mixer)
From 50 MHZ to 6,7 GHz 0 dBm
From 6,7 GHZ to 13,2 GHz - 3 dBm
From 13,2 GHz to 26,5 GHz - 5 dBm
Ranges of a conclusion to the screen Logarithmic scale 0.1, 0.2, 0.5 dB on division and from 1 to 20 dB into division at a step of 1 dB; 10 divisions on the screen
RBW ≥ 1 kHz From 0 to-85 dB from the calibrated basic level
The displayed average level of noise (DANL) (dBm) (the closed entrance, attenuation 0 dB, the detector of selections) 1 kHz of RBW; 30 Of RBW Hz VBW 10 Hz; 1 Of VBW Hz (Option - 1DR)
1 Of RBW Hz; 1 Of VBW Hz (Option - 1DR and - 1D5)25
KSVN on RCh an entrance (frequency is adjusted,
attenuation 10 dB)
100 Of Hz to 100 kHz 1,1:1 (Option - UKB)
9 kHz to 6,7ggts 1,3:1
6,7ggts to 13,2ggts 1,5:1
13,2ggts to 22,0ggts 2,0:1
22,0ggts to 26,5ggts 2,2:1
Inaccuracy of installation entrance attenyua-
Torah on 50 MHz

Attenuation
0 dB to 5 dB ±0,3дБ
10 dB Basic
15 dB ±0,3дБ
General amplitude accuracy ± (0,54db + absolute frequency characteristic)
Error of switching of width of a radio strip
(at the basic level)
1 kHz of RBW Basic
1 Of Hz to 3 Hz ±0,3дБ
10 kHz to 3 MHz of RBW ±0,3 dB
5 MHz of RBW ±0,6 dB
Range From - 149,9dbm to maximum
mixer level + installation of an attenuator
Permission Logarithmic scale ±0,1 dB
Linear scale ±0,12% of basic level
Accuracy (basic level - attenuation control + strengthening of a preamplifier) ±0,3 dB (-10 dBm to-60 dBm)
±0,5 dB (-60 dBm to-85 dBm)
±0,7 dB (-85 dBm to-90 dBm)

General characteristics of the analyzer of a range of Agilent E4407B:


Food
ON mode 195 … 250B of rms, 47 … 66 Hz
Power consumption <300 W
Power consumption, "Standby" <5 W
Alternating voltage 12 … 20B
Power consumption <200 W
Data storage Internal memory 8 Of MB
External devices 3,5" 1.44MB, MS-DOS diskette
Memory use Settings of the device 16kB
Settings + spectrogram (401 points) 20kB27
Speed of measurements Number of local measurements 40 amendments / with
The number of measurements through
GP-IB interface
40 amendments / with
Time of control central
frequencies
75 ms
Temperature range Worker
0 °C … +55 °C
Worker: 0 °C … +55 °C
For storage: -40 °C … +75 °C
Worker: 0 °C … +55 °C
For storage: -40 °C … +75 °C

Dimensions

Without handle 222 mm (V)) x 409 mm (D) x 373 mm (Sh)
With the handle (max) 222 of mm (In) x 516 mm (D) x 416 mm (Sh)

Information for the order:

Agilent Technologies E4407B Portable spectrum analyzer (CA) ESA-E 9 of kHz … 26,5ggts
Option-042/44 Gray/yellow backpack for transportation With / and
Option
Shipping method
Contact the seller
Analyzer of a range Agilent Technologies E4407B
Analyzer of a range Agilent Technologies E4407B
for
by the goods
Your message must contain at least 20 symbols. The message must not be more than 2000 symbols.
Obligatory field is not filled
The field is wrongly filled
The field is wrongly filled
Compare0
ClearSelected items: 0